Performance Test
4
Keysight U3401A User’s and Service Guide
77
Table 4-3
DC current verification test
Function
Input
Range
Error from nominal one year
DC current
0.00
μ
A
500
μ
A
±0.05
μ
A
0.0000 mA
5 mA
±0.4
μ
A
0.000 mA
50 mA
±4
μ
A
0.00 mA
500 mA
±40
μ
A
450.00
μ
A
500
μ
A
±0.27
μ
A
4.5000 mA
5 mA
±2.6
μ
A
45.000 mA
50 mA
±26
μ
A
450.00 mA
500 mA
±0.26 mA
0.0000 A
5 A
±0.5 mA
0.000 A
10 A
±5 mA
4.5000 A
5 A
±11.7 mA
9.000 A
10 A
±27 mA
CAUTION
Connect calibrator to multimeter’s A and Lo terminals before applying 10 A.