DC Power Supply Operation
Remote Sensing
120
U3606B User’s Guide
Enable remote sensing
To configure the U3606B for remote sensing:
1
Turn off the U3606B.
2
Remove the connections between the U3606B sensing (S+
and S–) and output (+ and –) terminals. Using a shielded
two- wire cable, connect the U3606B sensing terminals to
the load as shown in Figure 3- 1. Observe the polarity
when connecting the sensing leads to the load.
3
Turn on the U3606B.
4
Press
[Voltage]
to select the constant voltage mode. Use
the arrow keys to select the appropriate constant voltage
value.
5
Press
[Shift]
>
[EXT]
to enable remote sensing. When the
U3606B is operating in remote sensing mode, the EXT
annunciator on the front panel is illuminated.
6
Press
[
]
to regulate the output voltage.
7
Press
[Shift]
>
[EXT]
again to disable remote sensing when
the output state is on standby.
C A U T I O N
Do not use the shield as one of the sensing conductors. The other end
of the shielded two-wire cable should be left unconnected.
EXT annunciator
O U T
S B Y
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...