Operation and Features
2
Measuring Low-Resistance
U3606B User’s Guide
45
Table 2-6
Low-resistance test current values
[1]
[1] When the display shows “OL”, always apply the detection current.
Range
Detection
current
Test current
[2]
[2] Press
[Current]
to change the test current selected (see page 46).
Default
Range
Min
Max
100 m
Ω
0.015 A
1.0000 A
S2 (A)
1.0000 A
3.0000 A
1000 m
Ω
0.015 A
0.1000 A
S1 (A)
0.1000 A
0.3000 A
10
Ω
X
mA
[3]
[3] “
X
” follows the test current setting of 1000
Ω
range.
50.0 mA
S1 (mA)
50.0 mA
100.0 mA
100
Ω
X
10.0 mA
S1 (mA)
4.0 mA
30.0 mA
1000
Ω
X
6.0 mA
S1 (mA)
4.0 mA
10.0 mA