DC Power Supply Operation
Protection Functions
96
U3606B User’s Guide
N O T E
•
After adjusting the OV value, you can also press
[Limit]
again to save
the changes made or press
[Shift]
>
[Exit]
to discard the changes
made.
•
The OV feature is always enabled by default whenever you select
constant current mode. You cannot disable the OV feature.
•
If the OV value is set to a greater value than the OVP value, the OVP
value will be adjusted to equal the OV value.
•
If the OV value is set to zero, this will result in the output current being
dropped to zero for limiting.
•
The OV value is not applicable for square-wave output.
•
The OV value is limited by the range selected. Press
[Shift]
>
[Range]
to select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).
Table 3-3
CC mode over-voltage range and values
Range
Over-voltage value
Minimum
Maximum
S1 (30 V/1 A)
0 V
31.5 V
S1m (30 V/100 mA)
0 V
31.5 V
S2 (8 V/3 A)
0 V
8.4 V
S1S2 (autoranging)
N/A
[1]
[1] If the S1S2 (autoranging) range is selected, you will not be able to adjust the OV
value.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...