DC Power Supply Operation
3
Selecting a Range
U3606B User’s Guide
113
Selecting a Range
You can allow the U3606B to automatically select the range
using autoranging or you can select a fixed range using
manual ranging.
Key
Description
Press
[Shift]
>
[Range]
to toggle between the available ranges.
Table 3-6
Available ranges for DC power supply functions
Range
CV
[1]
[1] Available ranges affect sweep functions (ramp and scan), OCP, and OC values.
CC
[2]
[2] Available ranges affect sweep functions (ramp and scan), OVP, and OV values.
Square wave
S1 (30 V/1 A)
✔
✔
✔
S2 (8 V/3 A)
✔
✔
✔
S1m (30 V/100 mA)
-
✔
-
S2m (1000 mV/3 A)
✔
-
-
S1S2 (autoranging)
✔
✔
✔
N O T E
•
The S1 range is selected by default. You cannot change the range when
the output is enabled (OUT). The output should always be on stand-by
(SBY) before the range or output function can be changed.
•
The protection and limit values will always be set to the previous saved
value with respect to the range selected when the range is changed.