4
System-Related Operation
Utility Menu Items
138
U3606B User’s Guide
Configuring the ramp signal parameters
Press
[Voltage]
or
[Current]
if you wish to configure the scan
signal parameters for CV or CC output.
Select an appropriate output range by pressing
[Shift]
>
[Range]
. The ramp signal amplitude end position is limited
by the range selected.
Table 4-3
Ramp signal parameters
Ramp signal
Output
Items
Constant Voltage
Constant Current
Range
[1]
S1
S2
S2m
S1
S2
S1m
Amplitude end
position
[2]
0 V to
31.500 V
0 V to
8.400 V
0 V to
1050.0 mV
0 A to
1.0500 A
0 A to
3.1500 A
0 A to
105.00 mA
Number of steps
1 step to 10000 steps
Dwelling time
Fastest of the instrument output capability
[3]
[1] Range S1S2 (autoranging) is not supported for the Ramp function.
[2] Amplitude start position is fixed at 0 (V or A) by default.
[3] Typically ~100 ms per step.