Characteristics and Specifications
5
Digital Multimeter Specifications
U3606B User’s Guide
159
DC specifications
Table 5-1
DC accuracy specifications ± (% of r % of range)
Function
Range
[1]
Resolution
Test current
or burden
voltage
24 Hours
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperatur
e
coefficient
0 ºC to 18 ºC
28 ºC to 55
ºC
DC voltage
19.9999 mV
0.1
μ
V
-
0.012 + 0.04
0.015 + 0.04
0.025 + 0.04
0.0015
+ 0.0040
100.000 mV
1
μ
V
-
0.012 +
0.008
0.015 +
0.008
0.025 +
0.008
0.0015
+ 0.0008
1.00000 V
10
μ
V
-
0.012 +
0.005
0.015 +
0.005
0.025 +
0.005
0.0010
+ 0.0005
10.0000 V
100
μ
V
-
0.012 +
0.005
0.015 +
0.005
0.025 +
0.005
0.0020
+ 0.0005
100.000 V
1 mV
-
0.012 +
0.005
0.015 +
0.005
0.025 +
0.005
0.0015
+ 0.0005
1000.00 V
10 mV
-
0.012 +
0.005
0.015 +
0.005
0.025 +
0.005
0.0015
+ 0.0005
DC
current
[3]
10.0000 mA
0.1
μ
A
< 0.2 V
0.05 + 0.015
0.05 + 0.015
0.05 + 0.015
0.0060
+ 0.0005
100.000 mA
1
μ
A
< 0.2 V
0.05 + 0.005
0.05 + 0.005
0.05 + 0.005
0.0060
+ 0.0005
1.00000 A
10
μ
A
< 0.3 V
0.05 + 0.007
0.05 + 0.007
0.15 + 0.007
0.0100
+ 0.0005
3.0000 A
100
μ
A
< 0.7 V
0.05 + 0.007
0.05 + 0.007
0.15 + 0.007
0.0150
+ 0.0010
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...