Characteristics and Specifications
5
Digital Multimeter Specifications
U3606B User’s Guide
165
Frequency specifications
Table 5-3
Frequency accuracy specifications ± (% of r % of range)
Function
Range
Frequency range
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
Frequency
[1]
Voltage path:
100 mV to 750 V
< 2 Hz
0.18 + 0.003
0.005
< 20 Hz
0.04 + 0.003
0.005
20 Hz to 100 kHz
0.02 + 0.003
0.005
100 kHz to 300 kHz
0.02 + 0.003
0.005
Current path:
10 mA to 3 A
< 2 Hz
0.18 + 0.003
0.005
< 20 Hz
0.04 + 0.003
0.005
20 Hz to 10 kHz
0.02 + 0.003
0.005
[1] For 100 mV and 1 V ranges, the measurable frequency is up to 1 MHz at 0.5 V signal . Minimum input frequency is 1 Hz.
N O T E
All frequency counters are susceptible to errors when measuring low-voltage,
low-frequency signals. Shielding inputs from external noise pickup is critical for
minimizing measurement errors.