Operation and Features
2
Measuring Capacitance
U3606B User’s Guide
49
Capacitance measurement summary
N O T E
Measuring tips
•
The U3606B calculates capacitance by charging a capacitor with a
known current for a period of time and then measuring the voltage.
•
For measuring capacitance values greater than 10000
μ
F, discharge the
capacitor first, then select a suitable range for measurement. This will
speed up the measurement time and also ensure that the correct
capacitance value is obtained.
•
For measuring capacitance values greater than 1 mF, a delayed
response should be expected from the front panel display. For remote
interface operation, increase the SCPI query time out value. (Typically
> 10000 ms.)
•
For measuring small capacitance values, press
[Null]
with the test
leads open to subtract the residual capacitance of the instrument and
leads.
Table 2-7
Capacitance measurement summary
Item
Description
Available ranges
1 nF, 10 nF, 100 nF, 1
μ
F, 10
μ
F, 100
μ
F, 1000
μ
F, 10000
μ
F
Measurement
method
Computed from constant current source charge time,
typical 0.2 V to 1.4 V signal level
Input protection
1000 V
rms
on all ranges, < 0.3 A short circuit
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...