Operation and Features
2
Triggering the Multimeter
U3606B User’s Guide
77
Remote interface triggering
Immediate triggering
In the immediate trigger mode, the trigger signal is always
present. When you place the multimeter in the
wait- for- trigger state, the trigger is issued immediately. This
is the default trigger source for the U3606B.
Remote interface operation:
•
The
TRIGger:SOURce IMMediate
command selects the
immediate trigger source.
•
After selecting the trigger source, you must place the
instrument in the “wait- for- trigger” state using the
INITiate[:IMMediate]
or
READ?
command. A trigger
will not be accepted from the selected trigger source until
the instrument is in the “wait- for- trigger” state.
Refer to the
U3606B Programmer’s Reference
for the syntax
and complete descriptions for these commands.
Software (bus) triggering
The bus trigger mode corresponds in function as the Single
triggering mode, except the trigger is initiated by sending a
bus trigger command, after selecting
BUS
as the trigger
source.
N O T E
The immediate trigger mode is available from the remote interface only.
N O T E
The bus trigger mode is available from the remote interface only.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...