Step 2:

 Calculate number of turns 

required based on difference 

between factory set point and 

desired set point using the 

following chart.

Example:

HEPA filter with desired set point of 

0.9 in w.g.
Number of turns at 0.9 in w.g. = 4
Number of turns at factory set 

point of 0.6 in w.g. = 2.2

Number of turns to field set = 4 – 

2.2 = 1.8 (approximately)

NOTE: 

Maximum available pressure 

setting is 1.0” w.g. Do not exceed 4 

½ turns (total).

Step 3: 

Turn screw clockwise the 

number of turns calculated above. 

You will need a 1.5mm hex Allen 

wrench

Step 4: 

To check filter pressure 

set point and/or to fine tune

insert 

Magnehelic gauge into the static 

pressure port (located along knife 

edge of diffuser – diffuser face 

must be removed to access).

Step 5: 

Using an obstruction (piece 

of cardboard or equivalent), slowly 

increasing pressure on the unit by 

blocking off more and more of the 

diffuser face until the LED turns 

from green to yellow.

Step 6: 

Take pressure 

measurement at point the LED 

changed color and rotate the 

adjustment screw until desired 

pressure set point is reached.

Set P

oint in Inches w

.g.

Turns of Set Point Adjustment Screw

(from starting point at 0.25” w.g.)

1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
0.10
0.00

Table 2. Set Point Adjustment

Turns of Set Point Adjustment Screw

(from starting point at 0.25” w.g.)

S

et P

oint in inc

hes w

.g.

 1 

 2 

  3 

  4               5 

1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
0.10

0.00

0 1 2 3 4 5

RSR Filter LED Set Point Adjustment

For more information visit 

KeystoneCleanAir.com

v001

19

Kr

own 

Fan Filter Units | 

Installation Instructions

Summary of Contents for Krown

Page 1: ...Krown Fan Filter Units v001 Issue Date 02 23 22 2022 Keystone Clean Air Solutions All rights reserved INSTALLATION SERVICE MANUAL...

Page 2: ...pply Units 22 Balancing Ducted Supply Pressure Independent and Dependent Airflow Control 23 Balancing Ducted Supply Pressure Independent and Dependent Airflow Control 25 Service Instructions Pre Filte...

Page 3: ...the shipping company and file a freight claim Filter Information Ensure that all HEPA ULPA filters are stored in a clean dry location in a vertical position as labeled on the carton Do not open the H...

Page 4: ...e four 4 stickers located on the back of the plenum For ease of troubleshooting assistance replacement parts and general factory assistance it is recommended that the order number and model line strin...

Page 5: ...tring for Bench Top Replaceable filter Krown X X I SUPPLY 24 48 BTR HEPA FC ECM 115 CF TMB ECMSC TC DSW 115 ASSP WPF PL AL F EM PL WM F EM Step 3 Raise the unit through the ceiling grid at an angle un...

Page 6: ...ontroller Step 2 Wire speed controller according to wiring diagram Wall mounted speed controller will be wired in series with the motor Step 3 Attach the wall mounted controls faceplate to the electri...

Page 7: ...analog voltage signal from BAS others or Price WMSC please remove ECMSC from control box Connect 0 10VDC COMMON to BAS SIGNAL BAS COMMON on back of ECMSC make sure to feed wires through grommet on co...

Page 8: ...311 373 95 140 VDC CFM 19 207 VDC CFM 47 174 VDC CFM 70 197 VDC CFM 79 167 VDC CFM 138 212 VDC CFM 81 190 Supply Constant Torque Backward Curve Velocity 24x36 24x48 FPM BTR BC RSR BC BTR BC RSR BC 45...

Page 9: ...put BAS Voltage Response Notes 0 1 VDC Local control mode using push buttons Local setpoint can be adjusted from 0 100 using push buttons 1 2 VDC Motor Off Recommended sending a 1 5 VDC signal to comm...

Page 10: ...et Sample line string Krown X X I SUPPLY 24 48 RSR HEPA FC ECM 115 CF R TMB BACnet BFC TC DSW 115 PC 115 ASSP WPF PL AL F AL PL B12 F B12 Step 4 If the unit has an RSR filter install the filter now re...

Page 11: ...box The service tool connects to the RJ12 port on the BFC board to allow full control up to 6 feet away from the control panel Sample line string KrownCS X X I ECM BFC BFC ST 250002 600 For more infor...

Page 12: ...ECM CFM 5 08 44 ECM CFM 4 45 32 ECM CFM 6 93 42 ECM CFM 5 50 41 Supply Constant Torque Forward Curve Velocity 24x24 24x36 24x48 FPM BTR FC RSR FC BTR FC RSR FC BTR FC RSR FC 45 95 ECM CFM 10 82 12 ECM...

Page 13: ...quations are based on a 1 5 Volt scale The BAS input voltage is a 2 10 VDC scale and therefore VDC calculated for a given CFM using the equations must be doubled to achieve that CFM using the BAS inpu...

Page 14: ...gaurd Room Side Replaceable Filter RSR Quick Start Guide RSR BTR Bench Top Replaceable Filter BTR Sample line string Krown X X I 24 48 BTR FC ECM 115 10 CF HEPA BFC DSW 115 PC 115 TMB ASSP FL ML BACne...

Page 15: ...if the LED indicator option has been selected one port will be used for the LED installation ASSP on BTR unit ASSP INJ RSR unit Optional Aerosol Injection Port INJ Sample line string Krown X X I SUPPL...

Page 16: ...r Step 3 Remove threaded standoff and lock washer from LED cable assembly Step 4 Thread cable through static pressure port on face of BTR filter NOTE It is recommended to rest the filter frame on 2x4...

Page 17: ...nstallation section of this manual Step 8 Refer to wiring diagram located in control box to connect LED leads to terminal block NOTE If unit has BFC speed controller LED leads will need to be connecte...

Page 18: ...r operation when the static pressure is above 0 1 w g Red Troubleshoot Motor Indicates motor is not in operation and that unit static pressure is below 0 1 w g Calibration Filter LED will come factory...

Page 19: ...user diffuser face must be removed to access Step 5 Using an obstruction piece of cardboard or equivalent slowly increasing pressure on the unit by blocking off more and more of the diffuser face unti...

Page 20: ...lean filter pressure Red Off Adjust setpoint check control mode Indicates motor setpoint is 0 If alternate setpoint is expected please check control mode Blinking Red Low Pressure Troubleshoot motor I...

Page 21: ...ducted applications it is critical that primary air be balanced prior to installing the Krown Fan Filter unit The primary air controls should be locked in position post balancing and the is installed...

Page 22: ...Turn on Krown Fan Filter Unit and set to desired airflow See Airflow Setpoint section to determine initial setpoint Step 3 Measure airflow using flow hood and compare against scheduled flow Adjust sp...

Page 23: ...ee Airflow Setpoint section to determine initial setpoint Step 2 Confirm airflow using a flow hood and adjust speed controller as required Step 3 Ensure each supply air duct is balanced to scheduled a...

Page 24: ...Fan Filter Unit controller or upstream airflow device as required Repeat steps 4 6 until desired airflow and inlet pressure 0 0 0 05 w g is achieved FFU Airflow FFU Inlet Static measured Action Above...

Page 25: ...ed return proceed directly to Step 2 Step 2 Set Krown Fan Filter Unit to desired airflow See Airflow Setpoint section to determine initial setpoint Step 3 Confirm airflow using flow hood and adjust sp...

Page 26: ...out from under the brackets non ducted or remove filter access door ducted Reference the below figures for further details Step 3 Clean pre filter by hand washing in water with a mild detergent or by...

Page 27: ...ling or access panel Step 2 Remove filter by pushing filter to one side and sliding it out from under the brackets non ducted or remove the duct collar pre filter housing and pre filter if applicable...

Page 28: ...Step 6 Remove motor blower assembly from the inlet collar by removing the 6 lock nuts using a deep 7 16 socket wrench or equivalent Step 7 Inspect and or replace parts Step 8 Replace or reuse the mot...

Page 29: ...authorized service technician Step 1 Remove diffuser perforated face by turning quarter turn fasteners holding face in place NOTE Safety cables will be attached to the diffuser face but care should b...

Page 30: ...E A 7 16 ratchet with extension is recommended for ease of access for unfastening 2x3 and 2x4 forward curved blower housing 2x2 forward curved blower housing 2x3 and 2x4 backward curved blower housing...

Page 31: ...otor to control box Step 10 Remove motor blower assembly Step 11 Inspect and or replace parts Step 12 Replace with the new motor blower assembly and reassemble by reversing the above steps Step 13 Dou...

Page 32: ...by turning quarter turn fasteners holding face in place NOTE Safety cables will be attached to the diffuser face but care should be taken to ensure that diffuser face does not swing and injure anyone...

Page 33: ...ch the filter media as it can damage the filter If your unit is 24 x 48 or 24 x 36 skip to Step 8 Step 5 Remove the 2 nuts securing the baffle retainer bracket using a 7 16 socket wrench Location of N...

Page 34: ...the control lid Step 9 Remove the control lid from the inside of the plenum using a 7 16 socket wrench for access to the control box Step 10 Reverse the previous steps taken to re assemble the unit Ba...

Page 35: ...torque all that is required is to open the connection of the white wire as shown Switching from constant torque to constant flow To switch the motor from constant torque to constant flow all that is...

Page 36: ...ity procedure outlined above 4 Conduct smoke and photometer test on HEPA ULPA filter Seal or replace Motor Keeps Shutting Off 1 Check if unit is ducted 2 If no contact Price at criticalenvironments pr...

Page 37: ...ter 15 5x19 5x0 437 045964 004 MERV8 Pre filter 15 5x15 5x0 75 042313 055 MERV8 Pre filter 15 5x19 5x0 75 042313 043 Reverse Exhaust Flow PF 2x2 MERV8 Pre filter 19 5x19 5x0 75 042313 044 2x3 MERV8 Pr...

Page 38: ...Transformer 019436 004 208 240V 20VA Transformer 019436 013 277V 20VA Transformer 019436 007 Disconnect Switch Single Pole 15 Amp 277V 115V w o PC Option 019903 001 Dual Pole 30 Amp 208 240V 115V w PC...

Page 39: ...ocument contains the most current product information as of this printing For the most up to date product information please go to KeystoneCleanAir com 2022 Keystone Clean Air Solutions All rights res...

Reviews: