KFM2151
3-5
Operation
Scanned source
Scanning is performed in order from the smallest channel number. In the example of
Table 3-1, scanning progresses from left to right. Scanning is performed on channels
whose scanned source turned ON (enabled). Scanning is skipped for channels that
are turned OFF (disabled). In this example, CH3 is skipped and continues to CH4.
The VOLT/TERM display blinks.
V
al
u
e:
O
N
Scanned
OFF
N
ot scanned
Voltage protection (only for voltage scanning)
Set the overvoltage protection (OVP) and undervoltage protection (UVP).
V
al
u
e:
-2.00
V
to 200.00
V
Resol
u
tion:
W
hen
u
sing the
key
10 m
V
W
hen
u
sing SHIFT+
1
V
A buzzer sounds if the selectable range is exceeded.
Range setting (only for voltage scanning)
Set the range.
V
al
u
e:
2
V
, 20
V
, 200
V
, or AUTO
Ta
b
le 3-2
Meas
u
rement Range and A
u
to Range
Range
Meas
u
rement Range
A
u
to Range
Range Do
w
n
Range Up
200
V
-199.99 to 199.99
10
V
or less
N
one
20
V
-19.999 to 19.999
1
V
or less
20
V
or higher
2
V
-1.9999 to 1.9999
N
one
2
V
or higher
Ta
b
le 3-1
Scanning operation (example)
Logical channel n
u
m
b
er
CH1
CH2
CH3
CH4
Scanned so
u
rce
O
N
O
N
OFF
O
N
Scanning
Exec
u
ted
Exec
u
ted
Exec
u
ted
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......