3-14
KFM2151
Operation
3.9
Performing the Scanning Operation
The procedure to perform the impedance scanning and voltage scanning is
explained below.
3.9.1
Impedance Scanning
Scanning is performed on the channels on which scanning is enabled. The scanned
input signal is transmitted to the KFM2150 FC Impedance Meter. The impedance is
shown on the KFM2150.
There are two impedance scanning modes: manual scanning and auto scanning.
Manual scanning
Press the
and
keys or SHIFT+
and
keys in monitor mode to select the
channel number (1 to 32). The impedance is shown on the KFM2150.
Auto scanning
Scanning is performed on channels that are set to scanned source. Scanning is
performed in order from the smallest channel number. When the scanning of the last
channel is complete, the scanning operation returns to the smallest channel number.
This operation is repeated until the scanning operation is aborted.
If the measuring frequency on the KFM2150 is low, the scanning speed is slow.
You can switch to the voltage monitor by pressing the IMP/VOLT key during auto
scanning. Impedance scanning continues even while you monitor the voltage. You
cannot stop the impedance scanning while monitoring the voltage. If you want to do
so, press the IMP/VOLT key again and then stop it.
Scanning when using a single KFM2151
The procedure to scan when using a single KFM2151 is given below.
1.
Check that the KFM2151 is in monitor mode.
If the CHANNEL display or VOLT/TERM display is blinking, press the MON key.
2.
Press the IMP/
V
OLT key to select impedance.
The IMP LED illuminates. The monitor LED of the channel whose scanning is
enabled illuminates in green. The IMP LED and VOLT LED illuminate alternately
each time you press the IMP/VOLT key.
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......