KFM2151
3-15
Operation
3.
Press the SCA
N
key to start the scanning operation.
The SCAN LED illuminates while auto scanning is in progress. The minimum hold
time is fixed to 1 000 ms.
4.
Press the SCA
N
key to stop the scanning operation.
The scanning operation stops, and the SCAN LED turns off. You can also press the
and
keys or SHIFT+
and
keys to stop the scanning operation.
Scanning when the channels are expanded
The procedure to scan when using multiple KFM2151s is given below.
Scanning is performed on channels that are set to scanned source in order from the
smallest channel number. When the scanning on the master unit is complete, the
scanning starts on the slave units. When the scanning of the last channel is
complete, the scanning returns to the master unit. This operation is repeated until the
scanning operation is aborted.
1.
Check that the KFM2151 is in monitor mode.
If the CHANNEL display or VOLT/TERM display is blinking, press the MON key.
2.
Press the IMP/
V
OLT key to select impedance.
The IMP LED illuminates. The monitor LED of the channel whose scanning is
enabled illuminates in green. The IMP LED and VOLT LED illuminate alternately
each time you press the IMP/VOLT key.
3.
W
hile holding the SHIFT key do
w
n, press the SCA
N
key on the master
or sla
v
e
u
nit.
The SCAN LED of units that are scanning illuminates in orange.
The minimum hold time is fixed to 1 000 ms.
4.
Press the SCA
N
key on a
u
nit that is scanning to stop the scanning
operation.
The scanning operation stops, and the SCAN LED turns off.
To scan on a single unit, press the SCAN key on the desired unit.
3.9.2
Voltage scanning
Voltage scanning starts when you switch the KFM2151 to monitor mode after
setting the channels. There is no need to operate the SCAN key as in the impedance
scanning.
Voltage scanning stops while setting various conditions.
See
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......