Remote Control
4.4.2
Parameters
The parameter format of SCPI is derived from the program parameter format
defined in IEEE 488.2.
The representation system of the program data that the KFM 2151 handles is
indicated below.
Non-numeric parameters
■
String data
String data is used when a series of ASCII characters are requested.
Enclose a string in single quotations or double quotations. The start and end
quotation marks must match.
If you wish to use a quotation mark as a part of the string, enter two quotation marks
consecutively (with no characters in between).
■
Boolean data
Boolean data expresses a 1 or 0 condition or an ON or OFF condition. Responses
are returned as 1 or 0.
VOLTage:RANGe:AUTO ON
Numeric parameters
■
NR1
Represents an integer.
Details are given in the IEEE 488.2 Standard Digital Interface for Programmable
Instrumentation.
■
NR3
Represents a real n
u
m
b
er (exponential).
Details are given in the IEEE 488.2 Standard Digital Interface for Programmable
Instrumentation.
The value +5.20000+E01 is returned for the response data 52.0. The number of
digits to the right of the decimal is 5.
■
NRf
NRf is a generic term that includes NR1 and NR3.
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......