4-12
KFM2151
Remote Control
*IDN
Queries the manufacturer name, model, serial number, and firmware version.
Command
*IDN?
Response
The response to *IDN? is indicated below.
(Example)
For KFM2151 with a serial number AB123456 and firmware version 1.00
Returns KIKUSUI,KFM2151,AB123456,1.00.
•
When using a compound command in which multiple commands are
concatenated, include *IDN? as the last message unit. Otherwise, An SCPI error
(-440, "Query UNTERMINATED after indefinite response") occurs.
*OPC
The KFM2151 does not support asynchronous I/O operation.
When the *OPC command is transmitted, the Operation Complete bit (bit 0) of the
standard event status register is set.
Command
*OPC
*OPC?
Response
IEEE 4
88
.2-1992
Section 12.5.3
Returns the ACII character “1” in the output queue in response to *OPC?.
*RST
Resets the device to factory default settings. For the commands that are affected, see
Table 4-3.
Command
*RST
See