3-12
KFM2151
Operation
3.7
CONFIG Settings
This section explains how to set the CONFIG parameters for using the KFM2150.
The CONFIG parameter number and condition are displayed as shown in Fig.3-11.
Fig.3-11 Example of a CO
N
FIG setting display
Parameters and conditions
Ta
b
le 3-4
Parameter and settings
1.
Check that the KFM2151 is in monitor mode.
If the CHANNEL display or VOLT/TERM display is blinking, press the MON key.
2.
Press SHIFT+CO
N
FIG.
The CHANNEL display blinks, and parameter number 000 is shown.
3.
Press the
or
key to select the parameter n
u
m
b
er.
4.
Press the SET key.
The VOLT/TERM display blinks, and the conditions are shown.
5.
Press the
or
key to select the condition.
6.
Press the SET key to confirm.
The CHANNEL display blinks. To set other conditions, repeat steps step 3 and 6 .
7.
T
u
rn the PO
W
ER s
w
itch off and then t
u
rn it
b
ack on.
The specified conditions are activated.
Parameter
Nu
m
b
er
Parameter
Condition
000
RS232C
Ba
u
d rate
19 200
b
ps/9 600
b
ps/4
8
00
b
ps/2 400
b
ps
001
Data length
7
b
its/
8
b
its
002
Stop
b
it
1
b
it/2
b
its
003
Ackno
w
ledge message
O
N
/OFF
010
Unit n
u
m
b
er
00: Master
u
nit
01/02/03/04: Sla
v
e
u
nit
IMP
IMP
VOLT
VOLT
V
V
CH
CH
OVP
OVP
UVP
UVP
RANGE
RANGE
TERM
TERM
VOLT TERM
CHANNEL
Parameter number
Off
Off
Condition
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......