PAS SERIES
Maintenance 7-13
13.
Turn off the OUTPUT.
The voltage display shows "FS 3."
14.
Press the SET switch to store the calibration value.
The voltage display returns to "FS 0."
You are done calibrating the output voltage.
If you wish to exit from the output voltage calibration here, turn off
the POWER switch.
OVP (overvoltage protection) calibration procedure
Be sure to complete the calibration of the voltage system before per-
forming OVP calibration.
1.
2.
Connect a DVM to the output terminal.
■
Warming up
3.
Turn on the POWER switch while pressing the SET switch.
4.
Check that the current display shows "CAL."
If it is not, press the VOLTAGE switch.
5.
Warm up the equipment including the DVM for sufficient
time.
■
OVP offset and full scale
6.
Press the OVP switch to display "OVP" on the current display.
7.
Press the VOLTAGE switch to display "OF 0" on the voltage
display.
8.
Turn on the OUTPUT.
Calibration starts automatically (ON blinks) and the POWER
switch turns off. (This takes 30 s to 60 s.)
9.
Turn on the POWER switch while pressing the SET switch.
10.
Check that the current display shows "CAL."
If it is not, press the VOLTAGE switch.
11.
Press the OVP switch to display "OVP" on the current display.
12.
Press the VOLTAGE switch to display "FS 0" on the voltage
display.
13.
Calibration starts automatically (ON blinks) and the POWER
switch turns off. (This takes 30 s to 60 s.)
Summary of Contents for PAS 10-105
Page 10: ...VIII PAS SERIES This page has been intentionally left blank ...
Page 54: ...3 22 Basic Operation PAS SERIES This page has been intentionally left blank ...
Page 78: ...4 24 Remote Control PAS SERIES This page has been intentionally left blank ...
Page 131: ...PAS SERIES Specifications 8 13 This page has been intentionally left blank ...
Page 137: ...PAS SERIES Specifications 8 19 This page has been intentionally left blank ...
Page 145: ......