PAS SERIES
Specifications 8-5
Common Specifications
Notes
General (cont.)
Environment
specifications
Storage ambient
humidity
90 % RH or less.
No condensation.
—
Cooling system
Forced air cooling using a fan.
With thermal-sensing
control (FAN control)
Grounding polarity
Negative grounding or posi-
tive grounding possible.
—
Isolation voltage
± 500 Vmax (excluding the
500 V model)/± 600 Vmax
(500 V model)
—
Insulation
specifications
With-
standing
voltage
Across the
input terminals
and chassis
No abnormalities at
1 500 VAC for 1 minute.
—
Across the
input and out-
put terminals
No abnormalities at
1 500 VAC for 1 minute.
—
Across the out-
put terminals
and chassis
No abnormalities at 500 VDC
for 1 minute.
600 VDC for 1 minute
for 160 V, 320 V, and
500 V models.
Insula-
tion
resis-
tance
Across the
input terminals
and chassis
500 VDC, 30 M
Ω
or more.
At an ambient humidity of
70 % RH or less.
—
Across the
input and out-
put terminals
500 VDC, 30 M
Ω
or more.
At an ambient humidity of
70 % RH or less.
—
Across the out-
put terminals
and chassis
500 VDC, 30 M
Ω
or more.
At an ambient humidity of
70 % RH or less.
—
*1.
J1 connector on the rear panel. See the table 4-2 for the terminal arrangement.
*2.
Photocoupler open collector output. Maximum voltage 30 V, maximum current (sink) 8
mA. Insulated from the output and control circuits. Status signals are not mutually insulated.
*3.
If protection action is set to POWER switch off (BREAKER TRIP of CONFIG set to
Enable), the output (on) is held for approximately 0.5 s.
*4.
Output when the POWER switch is turned off manually, shut down using external analog
control, or turned off due to an ALM operation.
*5.
TP BUS connector on the rear panel.
*6.
The selection of (1) and (2) of EXT-V and EXT-R are made using CONFIG for both CV
and CC. The selected function is enabled for each.
The input impedance EXT-V CV CONT and EXT-V CC CONT is approximately 30 k
Ω
.
The setting error is ± 5 % of the rated output voltage (EXT-V CV CONT, EXT-R CV
CONT (1),(2)) and ± 5 % of the rated output current (EXT-V CC CONT, EXT-R CC
CONT (1), (2)).
*7.
OUTPUT off when BREAKER TRIP of CONFIG is set to Disable.
*8.
Under packaged condition.
Summary of Contents for PAS 10-105
Page 10: ...VIII PAS SERIES This page has been intentionally left blank ...
Page 54: ...3 22 Basic Operation PAS SERIES This page has been intentionally left blank ...
Page 78: ...4 24 Remote Control PAS SERIES This page has been intentionally left blank ...
Page 131: ...PAS SERIES Specifications 8 13 This page has been intentionally left blank ...
Page 137: ...PAS SERIES Specifications 8 19 This page has been intentionally left blank ...
Page 145: ......