4-14 Remote Control
PAS SERIES
4.1.5 Controlling the Output Current Using External
Resistance
This method is used to control the output current using an external
resistance (Rext) in the range 0 k
Ω
to approx. 10 k
Ω
.
Press the SET switch to check the preset current.
• The insulation of the external resistor (Rext) and the
• When using shielded wires for connection, protect
You can select the following two types of modes through CONFIG
settings. For details on CONFIG settings, see "3.2.4 Unit Configura-
tion (CONFIG)".
10 k
Ω
→
MAX OUT CC mode
The output current (Io) varies in the range of 0 to rated output cur-
rent by setting the external resistance (Rext) in the range of 0 k
Ω
to 10 k
Ω
.
Io = Irtg•Rext /10 [A]
Rext = 10•Io /Irtg [A]
Irtg: Rated output current [A]
10 k
Ω
→
0 OUT CC mode (FAIL SAFE)
The output current (Io) varies in the range of rated output current
to 0 by setting the external resistance (Rext) in the range of 0 k
Ω
to 10 k
Ω
.
Io = Irtg•(10-Rext)/10 [A]
Rext = 10•(Irtg-Io)/Irtg [A]
Irtg: Rated output current [A]
WARNING