Chapter 5 Maintenance
5.3
Troubleshooting
If you suspect that the PCR-M has malfunctioned, check whether any of the items
below apply to your case. In some cases, the problem can be solved quite easily.
If you find an item that corresponds to your case, follow the remedy for the item. If
you do not, we recommend that you initialize the PCR-M (memory contents are
cleared). If the remedy does not correct the problem, contact your Kikusui agent or
distributor.
■
The control panel display does not turn on even when the POWER switch is turned on.
■
A alarm sounds when the POWER switch is turned on.
■
The ALARM LED illuminates.
If your case does not correspond to any of the items above, see section 3.11,
“Protection Functions and Alarm Errors.”
See
Check Item
Check
Possible Cause
Remedy
Is the rated voltage being
applied to the AC inlet/
INPUT terminal block?
No
• Broken power cord
• Bad contact at the AC inlet of
the rear panel (PCR500M)
• Improper connection of the
power cord (PCR1000M /
PCR2000M)
Check that the power cord is not broken
and that the AC inlet connection is secure.
Yes
Malfunction
Remove the power plug from the outlet.
(PCR500M)
Turn off the circuit breaker of switchboard.
(PCR1000M / PCR2000M)
Immediately stop the use of the instrument
and request repairs.
Check Item
Check
Possible Cause
Remedy
Is the ALARM LED
illuminated or an error
number displayed?
Yes
An alarm or error
occurs.
Check the alarm or error type.
section 3.11, “Protection Functions and
No
Memory error
The PCR-M has been initialized due to memory
error. Memory contents have been cleared. If a
alarm sounds many times, request repairs.
Check Item
Check
Remedy
Is an alarm/error number
displayed?
Yes
An internal or external
error occurred on the
PCR-M.
Check the alarm type.
section 3.11, “Protection Functions and
Is the fan stopped?
Yes
The overheat protection
(alarm A-04) tripped
due to a fan failure.
Immediately stop the use of the instrument and
request repairs.
Is the air inlet or outlet
obstructed?
Yes
• The overheat
protection (alarm A-
04) tripped.
• Clogged dust filter.
Allow at least 20 cm between the air outlet and
the wall. Do not place objects within 20 cm.
Clean the dust filter.
Is the ambient temperature
exceeding 40 °C?
Yes
The overheat protection
(alarm A-04) tripped.
Use the PCR-M in an ambient temperature of
less than or equal to 40 °C. Move equipment that
generate heat at high temperatures away from
the PCR-M.