Chapter 2 Installation and Preparation
2.2
Precautions Concerning Installation Location
Be sure to observe the following precautions when installing the PCR-M.
●
Do not use the product in a flammable atmosphere.
To prevent the possibility of explosion or fire, do not use the product near alcohol,
thinner or other combustible materials, or in an atmosphere containing such vapors.
●
Avoid locations where the product is exposed to high temperature or
direct sunlight.
Do not install the product near a heater or in areas subject to drastic temperature
changes.
Operating temperature range: 0 °C to 40 °C (32 °F to 104 °F)
Storage temperature range:
-10 °C to 60 °C (14 °F to 140 °F)
●
Avoid humid environments.
Do not install the product in high-humidity locations--near a boiler, humidifier, or
water supply.
Operating humidity range:
20 %rh to 80 %rh (no condensation)
Storage humidity range:
90 %rh or less (no condensation)
Condensation may occur even within the operating humidity range. In such cases, do
not use the PCR-M until the condensation dries up completely.
●
Be sure to use it indoors.
This product is designed for safe indoor use.
●
Do not install the product in a corrosive atmosphere.
Do not install the product in a corrosive atmosphere or in environments containing
sulfuric acid mist, etc. This may cause corrosion of various conductors and bad
contacts of connectors inside the power supply leading to malfunction and failure, or
in the worst case, a fire.
●
Do not install the product in a dusty location.
Accumulation of dust can lead to electric shock or fire.
●
Do not use the product where ventilation is poor.
Secure adequate space around the product so that air can circulate around it.
Allow at least 20 cm of space between the air inlet/outlet and the wall (or obstacles).
●
Do not place objects on the PCR-M.
Placing objects on top of the product can cause failures (especially heavy objects).
●
Do not install the product on an inclined surface or location subject to
vibrations.
The product may fall or tip over causing damages and injuries.
●
Do not use the product in a location where strong magnetic or electric
fields are nearby or a location where large amount of distortion and noise
is present on the input power supply waveform.
The product may malfunction.
●
Do not use the foot on the side panel for standing the product in upright
position.
It may cause injury to the operator or the damage to the product when it falls down.