INIT:NAME
Starts the trigger function of the specified sequence group.
If the trigger source is set to IMM, the settings are changed immediately, or the
measurement, power line abnormality simulation, or sequence starts immediately. If the
trigger source is set to BUS, the PCR waits for a software trigger, and then the settings are
changed, or the measurement, power line abnormality simulation, or sequence starts.
Command
INITiate[:IMMediate]:NAME {TRANsient|ACQuire|SIMulation|PROGram}
Parameter
Value TRAN Starts the output change (sequence 1) trigger function
ACQ
Starts the measurement (sequence 3) trigger function
SIM
Starts the power line abnormality simulation trigger function (sequence
4; PCR-LE only)
This is not valid when the action that is performed when the current limit
).
PROG Starts the sequence operation trigger function (sequence 5; PCR-LE
only)
This is not valid when the action that is performed when the current limit
).
Introduction
Installing the VISA library
Interface Setup
Overview of Command
Parameters
Default settings
Command (function
search)
IEEE488.2 Common Commands
Output Setting
Protection and Clearing Alarms
Triggered Control of Output
Output On/Off Phase Control
Measured Value Queries
System Configuration
Registers
Command (ABC search)
Command (Sub-system
search)
SOURce
OUTPut
MEASure and FETCh
TRIGger
SYSTem
STATus
IEEE488.2 Common Commands
Other Commands
List(PDF)
Register
Architecture
Register structure
Appendix
A List of Errors
Processing time of main
Tutorial
Programming AC Output
Programming DC Output
Triggering Output Change
Output-on Phase Control
Measurement
Waiting for Operation Complete
Panel display
Status Monitoring
Error Checking
Visual Basic 2008
AC Power Supply PCR-M series
Communication Interface Manual
PCR-M series Communication Interface Manual
KIKUSUI ELECTRONICS CORP.
127/166