MEAS:CURR:AMPL:MAX:HOLD/
FETC:CURR:AMPL:MAX:HOLD
MEAS:CURR:AMPL:MAX:HOLD? starts a new measurement and then queries the peak
current (the held value).
FETC:CURR:AMPL:MAX:HOLD? queries the peak current (the held value) without starting a
new measurement.
You can use
to clear the held peak value.
command) are valid.
Command
MEASure[0][:SCALar]:CURRent:AMPLitude:MAXimum:HOLD?
FETCh[0][:SCALar]:CURRent:AMPLitude:MAXimumv:HOLD?
Response
Returns the peak current (the held value) in NR3 format in response to
MEAS:CURR:AMPL:MAX:HOLD? or FETC:CURR:AMPL:MAX:HOLD?.
In case of the single-phase three-wire output, MEAS:CURR:AMPL:MAX:HOLD? or
FETC:CURR:AMPL:MAX:HOLD? returns a comma-separated list of the peak current (the held
value) <NR3> for U and V phases.
In case of the three-phase output, MEAS:CURR:AMPL:MAX:HOLD? or
FETC:CURR:AMPL:MAX:HOLD? returns a comma-separated list of the peak current (the held
value) <NR3> for U, V, and W phases.
Unit
:
A
Introduction
Installing the VISA library
Interface Setup
Overview of Command
Parameters
Default settings
Command (function
search)
IEEE488.2 Common Commands
Output Setting
Protection and Clearing Alarms
Triggered Control of Output
Output On/Off Phase Control
Measured Value Queries
System Configuration
Registers
Command (ABC search)
Command (Sub-system
search)
SOURce
OUTPut
MEASure and FETCh
TRIGger
SYSTem
STATus
IEEE488.2 Common Commands
Other Commands
List(PDF)
Register
Architecture
Register structure
Appendix
A List of Errors
Processing time of main
Tutorial
Programming AC Output
Programming DC Output
Triggering Output Change
Output-on Phase Control
Measurement
Waiting for Operation Complete
Panel display
Status Monitoring
Error Checking
Visual Basic 2008
AC Power Supply PCR-M series
Communication Interface Manual
PCR-M series Communication Interface Manual
KIKUSUI ELECTRONICS CORP.
134/166