CURR:PROT:STAT
Sets the PCR acts when the current limit is exceeded.
You cannot set the time (it is fixed to 3 seconds).
Command
[SOURce:]CURRent:PROTection:STATe {ON|OFF|1|0}
[SOURce:]CURRent:PROTection:STATe?
Parameter
Value ON(1) TRIP (when an overload occurs for longer than the specified time, the
output is turned off, and an alarm is generated; default).
OFF(0) CC (when an overload occurs, the output is lowered so that the current
limit is not exceeded).
For the setting that is applied when *RST is sent, see this
Response
Response
Returns the PCR acts when the current limit is exceeded in NR1 format in response to
CURR:PROT:STAT?.
Introduction
Installing the VISA library
Interface Setup
Overview of Command
Parameters
Default settings
Command (function
search)
IEEE488.2 Common Commands
Output Setting
Protection and Clearing Alarms
Triggered Control of Output
Output On/Off Phase Control
Measured Value Queries
System Configuration
Registers
Command (ABC search)
Command (Sub-system
search)
SOURce
OUTPut
MEASure and FETCh
TRIGger
SYSTem
STATus
IEEE488.2 Common Commands
Other Commands
List(PDF)
Register
Architecture
Register structure
Appendix
A List of Errors
Processing time of main
Tutorial
Programming AC Output
Programming DC Output
Triggering Output Change
Output-on Phase Control
Measurement
Waiting for Operation Complete
Panel display
Status Monitoring
Error Checking
Visual Basic 2008
AC Power Supply PCR-M series
Communication Interface Manual
PCR-M series Communication Interface Manual
KIKUSUI ELECTRONICS CORP.
85/166