Setting the Electronic Load
Setting the protection functions
If the unit of the PLZ-4W series is set as the factory default settings, it is not necessary to set the
protection function.
Especially, when the DUT is concerned the safety as a matter of high priority or the charge and
discharge test is conducted with a risk, the level of safety can be improved by the setting of
protection functions.
•
Over Current protection (OCP) : 105 % of the maximum discharge current
When the OCP or the others alarm of the PLZ-4W series activates while the charge and discharge
testing is executed, it becomes the state of the CD/B alarm by the PFX2512 and the load is turned
off. Press the ENTER key of the PLZ-4W series, and remove the cause of alarm occurrence. Then after
releasing the alarm of the PLZ-4W series, release the alarm from the application software
BPChecker3000.
For details of setting procedure such as setting and releasing the alarm when the protection
function is activated, please refer to the operation manual of the PLZ-4W series, "Basic operation".
Specifing the master unit and slave units of the parallel
operation
1
By the setting menu concerning to the master slave parallel operation, set the
following conditions respectively.
•Master unit Menu > Configuration > Master/Slave > Operation : Master
•Slave unit Menu > Configuration > Master/Slave > Operation : Slave
2
On the master unit, specify the connecting number of slave units.
•Two units Menu > Configuration > Master/Slave > Parallel
:
2
For detailed setting procedure, refer to the chapter "Basic operation" and "Applied operation" of the
operation manual of PLZ-4W series.
See
Summary of Contents for PFX2515
Page 10: ...10 PFX2512_CE This page is intentionally blank ...
Page 58: ...58 PFX2512_CE This page is intentionally blank ...
Page 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Page 86: ...86 PFX2512_CE This page is intentionally blank ...
Page 100: ...100 PFX2512_CE This page is intentionally blank ...
Page 103: ......