Making the DUT cable
1
Refer to the following table, select the cable.
2
Attach the crimp terminal of M6 to the cable end connecting to the PFX2512.
■
Nominal cross-sectional area of cables and allowable currents (reference)
CAUTION
When the DUT is exposed under the high temperature such as using a temperature chamber,
please pay attention for the allowable temperature of the DUT cable. If the temperature of heat
resistant of the DUT cable is not sufficient, the DUT may get into the condition of danger caused by
the insufficient insulation and increasing the contact resistance.
• The length of the cable can be extended up to 5 m. If any longer length of the cable is used, the
constant current control may become unstable and it may effect to the accurate capacity
measurement.
• To prevent erroneous wiring, change a color of the cable or the crimp terminal to verify the
positive (+) pole and the negative (-) pole easily.
Nominal cross-
sectional area
[mm
2
]
AWG
(Reference cross-
sectional area)
[mm
2
]
Allowable
current
*1
[A] (Ta = 30
°
C)
*1.
Excerpts from Japanese laws related to electrical equipment.
Current
recommended
by Kikusui [A]
2
14
(
2.08
)
27
10
3.5
12
(
3.31
)
37
-
5.5
10
(
5.26
)
49
20
8
8
(
8.37
)
61
30
14
6
(
13.3
)
88
50
22
4
(
21.15
)
115
80
30
2
(
33.62
)
139
-
38
1
(
42.41
)
162
100
50
1/0
(
53.49
)
190
-
60
2/0
(
67.43
)
217
-
80
3/0
(
85.01
)
257
200
100
4/0
(
107.2
)
298
-
125
-
-
344
-
150
-
-
395
300
200
-
-
469
-
Summary of Contents for PFX2515
Page 10: ...10 PFX2512_CE This page is intentionally blank ...
Page 58: ...58 PFX2512_CE This page is intentionally blank ...
Page 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Page 86: ...86 PFX2512_CE This page is intentionally blank ...
Page 100: ...100 PFX2512_CE This page is intentionally blank ...
Page 103: ......