54
PFX2512_CE
Protection Functions and Alarms
When the protection function interrupts or the alarm occurs while the test, either turning off the
output or illuminate the "ALARM/WARNING" LED (in red or orange) and interrupts testing.
In case the PC operation stopped suddenly, the charge and discharge test is executed for the cycle
of condition just prior to the occurrence, but not go to the next cycle and quit halfway of the
testing. And the test data can not be saved.
For emergency, entire charge / discharge test can be stopped immediately.
Refer to the help file of the BPChecker3000 for details of the setting procedure and the counter
measure for protection functions, confirming status of the equipments and the system, repeating
of the test.
Alarm occurrence
Test Condition Editor
Test Executive
PWR series
PLZ-4W series
The PFX2512 equips the following protection features.
When either receiving the alarm, or detecting the error, or
activating the protection function, the "ALARM/WARNING" LED
illuminates in red, and the error code is displayed on the front
panel. The error code display can be back to the normal display
by pressing the SELECT key or by releasing the alarm from the
BPChecker3000 .
The delay (delaying the operation) can be set for the S_OVP,
S_UVP, and S_OCP.
?
Help
Ope
Example when the detection of wiring
error is occurred
Illuminate (Red)
Panel display
Cause of alarms
・
Descriptions
Operating
status
Err.001
Conn
Detection of abnormal wirings, DUT connection error (
Alarm LED
illuminate
Err.002
S_OVP
Software over voltage (over charge) protection
Alarm LED
illuminate
Output off
Err.003
S_UVP
Software low voltage (over discharge) protection
Err.004
S_OCP
Software over current protection
Err.005
OTP
Over temperature / DUT (battery) over temperature protection
Err.006
OAH
Over charge
・
over charge capacity / over discharge capacity protection
Err.007
Comm
Communication error detection, command alarm error
Alarm LED
illuminate
Err.008
AC_OFF
Abnormal AC line detection, Power line alarm
Alarm LED
illuminate
Output off
Err.009
H_OVP
Hardware over voltage (over charge) protection
Err.010
H_UVP
Hardware low voltage (over discharge) protection
Err.011
H_OCP
Hardware over current protection
Err.012
OHP
Over heat protection
Err.013
Shock_Det.
Vibration alarm
Err.014
EXT_ALM
External trigger alarm
Err.050
CD/B
PLZ_ALM
Electronic load alarm
Err.051
PLZ_CONT Exceeding control voltage of the Electronic load
See
Summary of Contents for PFX2515
Page 10: ...10 PFX2512_CE This page is intentionally blank ...
Page 58: ...58 PFX2512_CE This page is intentionally blank ...
Page 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Page 86: ...86 PFX2512_CE This page is intentionally blank ...
Page 100: ...100 PFX2512_CE This page is intentionally blank ...
Page 103: ......