10 PLZ2004WB/PLZ2004WHB

 

General Specifications

PLZ2004WB

PLZ2004WHB

Environment Operating 

temperature range

 0 ºC to 40 ºC (32 ºF to 104 ºF)

Operating humidity 
range

20 %rh to 85 %rh (no condensation)

Storage 
temperature  range

-25 ºC to 70 ºC (-13 ºF to 158 ºF)

Storage humidity 
range

Less than or equal to 90 %rh  (no 
condensation)

Input line voltage range

100 Vac to 240 Vac
(90 Vac to 250 Vac) single phase, continuous

Input frequency range

47 Hz to 63 Hz

Power consumption

200 VAmax

Inrush current

35 A

120 A

1

Protective conductor current

---

600 µA (100 Vac at 
50 Hz; TYP)

Insulation 
resistance

Primary - input 
connector, 
Primary - chassis, 
Input connector - 
chassis

500 Vdc,
30 MΩ or more
(ambient humidity of 
70 %rh or less)

1 000 Vdc,
30 MΩ or more
(ambient humidity of 
70 %rh or less)

Withstand 
voltage

Primary - input 
connector, 
Primary - chassis,

No abnormalities at 1 500 Vac for 1 minute

Dimensions

See outline drawing.

Weight

Approx. 24 kg (52.91 lb)

Accessories Power cord

1 pc. (Cable length of 2.4 m)

Load input terminal 
cover

2 sets (cover and 
auxiliary band)

1 pc. (cover and two 
lock plates)

Screw set for th 
load input

2 sets (bolts, nuts, and spring washers)

Heavy object 
warning label

1 pc.

Operation manual

1 pc.

Electromagnetic compatibility
(EMC)

3 4

Conforms to the requirements of the 
following directive and standard.
EMC Directive 2014/30/EU
 EN 61326-1
 EN 61000-3-2
 EN 61000-3-3

Safety

 3

Conforms to the requirements of the 
following directive and standard.
Low Voltage Directive 2014/35/EU

4

EN 61010-1 (Class I

5

, Pollution degree 2

6

)

1. 

Approximately 70 A when 100 Vac is applied

2. 

If the input voltage or input frequency is different, the following equation can be 
used to calculate the value.

 

               Input voltage [V]   Input frequency [Hz]

Protective conductor current =                            ×   

     

           × 600 [µA]

 

                     100 [V] 

  

     50 [Hz]

3. 

Not applicable to custom order models.

4. 

Only on models that have CE marking on the panel.

5. 

This instrument is a Class I equipment. Be sure to ground the protective conductor 
terminal of the instrument. The safety of the instrument is not guaranteed unless 
the instrument is grounded properly

6. 

Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a 
reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes 
that only non-conductive pollution will occur except for an occasional temporary 
conductivity caused by condensation.

Dimensions

PLZ2004WB 

Unit: mm (inch)

MAX150  (5.91)

128 (5.04)

429.5 (16.91)

MAX440 (17.32)

550 (21.65)

MAX600 (23.62)

430 (16.93)

173 (6.81)

MAX190 (7.48)

550 (21.65)

MAX590 (23.23)

PLZ2004WHB 

Unit: mm (inch)

Trademarks

Microsoft and Windows are either registered trademarks or 
trademarks of Microsoft Corporation in the United States and/or other 
countries.
Other company names and product names used in this manual are 
generally trademarks or registered trademarks of the respective 
companies.

Inquiry about Product

When making an inquiry about the product, please provide us with 
the following information.

Model (indicated at the top section on the front panel)
Serial number (indicated at the bottom section on the rear panel)

Copyrights

Reproducing or reprinting this operation manual or any part of it 
without the permission of the copyright holder is strictly prohibited.
Product specifications and manual contents are subject to change 
without notice.

© 2010 Kikusui Electronics Corp.

10 

环境保护使用期限

Environment-friendly Use Period

该标记为适用于在中华人民共和国销售的电子信息产品的环境保
护使用期限。
只要遵守有关该产品的安全及使用注意事项,从制造年月起计算,
在该年度内,就不会对环境污染、人身、财产产生重大的影响。
产品的废弃请遵守有关规定。
产品的制造年月可以在以下网址中确认。

     

http://www.kikusui.co.jp/pi/

有毒有害物质或元素名称及含有標示

Name of hazardous materials and symbol of element in the 

equipment and quantity

部件名称

有毒有害物质或元素

Pb

Hg

Cd

六价铬

Cr(VI)

多溴联

PBB

多溴二

苯醚

PBDE

印刷电路板组装品

×

×

内部接线

×

外壳

×

底盘组装品

×

×

辅助设备

×

本表格依据

SJ/T 11364

的规定编制。

: 该部件所有均质材料的有毒有害物质的含量不超过

GB/T 26572

准所规定的极限值要求。

×

: 该部件至少有一种均质材料的有毒有害物质的含量超过

GB/T 26572

标准所规定的极限值要求。

Summary of Contents for PLZ2004WB

Page 1: ...wattage The units connected in parallel can be used as a single electronic load The connection of the control cables is easy The control cable used to connect between the master unit and the load booster and between each load booster is one flat cable each There is no power switch The AC input power is turned ON OFF by the master unit Checking the Package Contents When you receive the product chec...

Page 2: ...dge This is to prevent the possibility of personal injury Purpose Never use the product for purposes other than the productʼs intended use This product is not designed or manufactured for general home or consumer use Input power Line Voltage Use the product within the rated input power voltage range For applying power use the power cord provided see page 7 This product is designed as an equipment ...

Page 3: ...netic or electric fields are nearby or a location where large amount of distortion and noise is present on the input power supply waveform The product may malfunction Do not stack more than two units on top of each other The units load booster and master unit can be stacked but do not stack more than two units on top of each other for safety reasons If you are using multiple load boosters it is rec...

Page 4: ...ut terminal Input connector used to connect to the equipment under test 7 AC INPUT connector AC power input connector 8 Serial number 9 Air outlet Air outlet used to exhaust the internal heat using a fan POWER ELECTRONIC LOAD 2000W x xxxV 0 xxxA PLZ2004WB DC INPUT 2000W 1 5 150V 0 400A AC INPUT 100 240V 47 63Hz 200VA MAX J1 J2 PROTECTIVE GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND DO NOT REMOV...

Page 5: ... booster Continue connecting all the remaining boosters in the same manner This completes the procedure for connecting the signal wires Load wiring The cautions and things that you need to consider when you are wiring the loads are the same as when you are wiring the loads of the PLZ1004W PLZ1004WH To ensure operational stability see Methods for Ensuring Stable Operation in chapter 1 Installation ...

Page 6: ... input terminals on the booster using the included load input terminal screw set Bolt M12 25 Spring washer M12 Nut M12 Crimping terminal Be sure to use the included screw set 6 For each of the booster s load input terminals wrap the included auxiliary band around the space between the rear panel and the bolt The auxiliary band prevents the load input terminal cover from slipping Wrap the belt tigh...

Page 7: ... Unit After you finish making the connections set the number of boosters on the master unit If the master unit s response speed has been set to 1 1 its response speed is changed to 1 2 to ensure operational stability You can use the menu settings to return the response speed to 1 1 1 Turn on the power to the master unit On the master unit from the menu select 2 Configuration 1 Master Slave Operati...

Page 8: ...e Pin Cleaning the PLZ2004WHB s dust filter 1 Remove the lower louver from the panel While lifting the bottom of the removal mark with your finger tips slide the entire louver to the right Then pull it toward you POWER ELECTRONIC LOAD PLZ2004WHB 2000W 5 650V 0 100A While lifting the bottom of the removal mark with your finger tips slide the entire louver to the right Removal mark Louver 2 Remove al...

Page 9: ...ally increased from 0 V no current will flow until 0 3 V is exceeded Once a current greater than or equal to 1 of the range rating starts flowing the current can flow at voltages less than equal to 0 3 V 2 Condition in which the booster is connected to the master unit 3 M range applies to the full scale of H range Rating PLZ2004WHB PLZ2004WHB Operating voltage 5 Vdc to 650 Vdc1 Current 100 A Power...

Page 10: ... on models that have CE marking on the panel 5 This instrument is a Class I equipment Be sure to ground the protective conductor terminal of the instrument The safety of the instrument is not guaranteed unless the instrument is grounded properly 6 Pollution is addition of foreign matter solid liquid or gaseous that may produce a reduction of dielectric strength or surface resistivity Pollution Deg...

Reviews: