PWR-01
21
Installation and Preparation
|
Selecting the Load Cables
Limitations of the remote sensing function
All wires have resistance. As the wire becomes longer or the current becomes larger, the volt-
age drop in the wire becomes greater. This results in a smaller voltage being applied at the
load end. The PWR-01 has a sensing function that compensates for this voltage drop up to
the following values
. If the voltage drop exceeds this level, use cables that have a
greater cross-sectional area.
L type
ML type
MH type
H type
Compensation
voltage
Approx. 1.5 V one
way
Approx. 4 V one
way
Approx. 5 V one
way
Approx. 5 V one
way
Summary of Contents for PWR1201H