PWR-01
107
Parallel/ Series Operation
|
Master-Slave Parallel Operation
Starting master-slave parallel operation
Turning on the POWER switch
1
Turn on the master unit’s POWER switch.
2
Turn off the POWER switch of each slave unit.
3
Operate the master unit normally.
You cannot perform panel operations on the slave units. Turn output on and off from
the master unit.
Turning the POWER switch off
1
Turn off the POWER switch of each slave unit.
2
Turn off the master unit’s POWER switch.
CAUTION
After you turn the POWER switch off, wait at least 10 seconds after the panel display turns
off before you turn the POWER switch back on. Repeatedly turning the POWER switch on
and off at short intervals can cause damage to the inrush current limiter. Furthermore, this
will shorten the service life of the POWER switch, internal input fuse, and other components.