Basic Features
|
Protection functions and Alarms
Front-panel output terminal overcurrent protection (FOCP)
Front-panel output terminal overcurrent protection (FOCP) is activated when the output cur-
rent from the front-panel output terminal exceeds 10 A (typical value). The trip point is fixed.
If the FOCP trip point is lower than the OCP trip point, OCP is prioritized.
Overheat protection (OHP)
Overheat protection (OHP) is activated when the PWR-01’s internal temperature rises to an
abnormal level.
• When the PWR-01 is used in an environment that exceeds its operating ambient tem-
perature range (0 °C to +50 °C).
• When the PWR-01 is used with its air inlet or outlet port blocked.
• When the fan motor has stopped.
If you do not fix the problem that caused the OHP to be activated, OHP will be activated again
even if you turn the POWER switch off and on (
p.61
).
Incorrect sensing connection protection (SENSE)
Incorrect sensing connection protection is activated when the remote sensing cables are con-
nected with incorrect polarity (positive or negative), and the remote sensing function is then
used.
Low AC input protection (AC-FAIL)
Low AC input protection is activated when a low AC power line input voltage is detected.
You can set the PWR-01 so that when the problem that caused the AC-FAIL to be activated is
corrected, the output is automatically turned on (CF20:
AUTO
).
Activation and release are repeated according to the variation in the AC power line.
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
B
C
A
PRESET
DLY
SEQ
HB
SS
VIR
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
SC2
SC3
LOCK
ALM CLR
SC1
SHIF T
FINE
FINE
REGULATED DC POWER SUPPLY
OUTPUT
VOLTAGE
CURRENT
CONFIG
PWR DSPL
MEMORY
LOCAL
SET
OVP•OCP
Lit
Blinking
(orange)
Alarm indication when FOCP is activated
Alarm indication when OHP is activated
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
B
C
A
PRESET
DLY
SEQ
HB
SS
VIR
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
SC2
SC3
LOCK
ALM CLR
SC1
SHIF T
FINE
FINE
REGULATED DC POWER SUPPLY
OUTPUT
VOLTAGE
CURRENT
CONFIG
PWR DSPL
MEMORY
LOCAL
SET
OVP•OCP
Lit
Alarm indication when incorrect sensing connection
protection (SENSE) has been activated
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
B
C
A
PRESET
DLY
SEQ
HB
SS
VIR
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
SC2
SC3
LOCK
ALM CLR
SC1
SHIF T
FINE
FINE
REGULATED DC POWER SUPPLY
OUTPUT
VOLTAGE
CURRENT
CONFIG
PWR DSPL
MEMORY
LOCAL
SET
OVP•OCP
Lit
Alarm indication when low AC input protection
(AC-FAIL) is activated
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
B
C
A
PRESET
DLY
SEQ
HB
SS
VIR
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
SC2
SC3
LOCK
ALM CLR
SC1
SHIF T
FINE
FINE
REGULATED DC POWER SUPPLY
OUTPUT
VOLTAGE
CURRENT
CONFIG
PWR DSPL
MEMORY
LOCAL
SET
OVP•OCP
Lit
Summary of Contents for PWR1201H
Page 7: ...PWR 01 7 This page is intentionally blank ...
Page 12: ...12 PWR 01 This page is intentionally blank ...
Page 74: ...74 PWR 01 This page is intentionally blank ...
Page 100: ...100 PWR 01 This page is intentionally blank ...
Page 112: ...112 PWR 01 This page is intentionally blank ...
Page 113: ...Maintenance This chapter explains how to perform cali bration ...
Page 134: ...134 PWR 01 This page is intentionally blank ...
Page 135: ...Appendix A Factory Default Settings B Options C Troubleshooting ...
Page 144: ...144 PWR 01 ...