52
PWR-01
Basic Features
|
Protection functions and Alarms
Setting limits
Voltage setting limit
The maximum voltage that can be set is limited to about 95 % of the OVP trip point to prevent
OVP activation due to mistaken operations.
By factory default, the voltage setting limit is set to on (CF23:
ON
).
The voltage setting limit (CF23) on/ off state is synchronized to undervoltage limit (UVL).
If you enable the voltage setting limit when the voltage setting is higher than 95 % of the OVP
trip point, the voltage setting is retained, but the OVP trip point is changed to 105 % of the
voltage setting.
Undervoltage limit (UVL)
Undervoltage limit (UVL) limits the minimum voltage that can be set to the UVL trip point.
Undervoltage limit (UVL) is synchronized to the on/off state of voltage setting limit (CF23).
By factory default, the voltage setting limit is set to on (CF23:
ON
).
If you enable the voltage setting limit when the voltage setting is less than the set UVL trip
point, the voltage setting is retained, but the UVL trip point is changed to the voltage setting.
Setting the UVL trip point
1
Press OVP•OCP twice.
The present UVL trip point is displayed.
2
Turn the VOLTAGE knob to set the UVL trip point (
UVL setting range: 0 % to 105 % of the rated output voltage
3
Press OVP•OCP to finish the configuration.
The UVL trip point will be set. The OVP•OCP key turns off, and the measured value
display appears.
L type
ML type
MH type
H type
UVL setting range
0 V to 42 V
0 V to 84 V
0 V to 252 V
0 V to 682.5 V
Display example of the UVL setting
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
B
C
A
PRESET
DLY
SEQ
HB
SS
VIR
CV
CC
/W
V
A
ALM
RMT
LOCK
LAN
SC2
SC3
LOCK
ALM CLR
SC1
SHIF T
FINE
FINE
REGULATED DC POWER SUPPLY
OUTPUT
VOLTAGE
CURRENT
CONFIG
PWR DSPL
MEMORY
LOCAL
SET
OVP•OCP
Lit
Summary of Contents for PWR1201H
Page 7: ...PWR 01 7 This page is intentionally blank ...
Page 12: ...12 PWR 01 This page is intentionally blank ...
Page 74: ...74 PWR 01 This page is intentionally blank ...
Page 100: ...100 PWR 01 This page is intentionally blank ...
Page 112: ...112 PWR 01 This page is intentionally blank ...
Page 113: ...Maintenance This chapter explains how to perform cali bration ...
Page 134: ...134 PWR 01 This page is intentionally blank ...
Page 135: ...Appendix A Factory Default Settings B Options C Troubleshooting ...