5-52
TOS9213S/ TOS9213AS
5.11 Program
By combining test conditions stored in internal memory, up to 100 programs com-
posed of a total of 500 steps can be created.
Each program can accommodate 100 steps.
For details on memory, see "5.10 Panel Memory".
Press the AUTO key. The LED lights up and the program screen (AUTO) is dis-
played.
Example of a program
To create the above program, make the following settings:
00: DCW01-HOLD
01: IR01-0.5s
END
(Explanation of the program example)
At Step 00, DCW01 (DC withstanding voltage test) is conducted. The Step-00 inter-
val time is set to HOLD. Thus, Step 01 does not start until the START switch is
pressed. When START is pressed, IR01 (insulation resistance test) ends at Step 01
and, 0.5 seconds later, the tester enters the READY status.
When END is set to RET, the test restarts at Step 00.
AUTO
AUTO
READY
PRG
00 : --UNTITLED--
TOTAL: 2
01 DCW02-HOLD
02 IR 03-0.5s
03
Program name
Step No.
Program No.
Test step
Total number of steps
Repeat-test setting
END
Memory
DCW01
Interval
HOLD
Memory
IR01
Interval
0.5 s
Step 00
Step 01
Flow of a program test