TOS9213S/ TOS9213AS
6-3
Control procedure
1.
T
u
rn off the PO
W
ER s
w
itch.
2.
Using the excl
u
si
v
e connection ca
b
le (5-pin DI
N
ca
b
le), connect the
optional de
v
ice to the REMOTE connector on the front panel.
3.
T
u
rn on the PO
W
ER s
w
itch.
A test can be started on the optional device. In such a case, the START switch
on the panel becomes invalid. To stop the test, either press the STOP switch on
the panel or input the STOP signal on the optional device. For details, see the
operation manual for the optional device.
4.
To ret
u
rn to the panel for control, first t
u
rn off the PO
W
ER s
w
itch on the
tester.
5.
On the front panel, remo
v
e the excl
u
si
v
e connection ca
b
le (5-pin DI
N
ca
b
le) from the REMOTE connector.
6.
T
u
rn on the PO
W
ER s
w
itch.
The START switch on the panel becomes valid.
• When the START signal of the SIGNAL I/O connector is valid, the signal has pri-
ority over the START signal of the REMOTE terminal. The STOP signals of the
SIGNAL I/O connector and the REMOTE terminal are accepted equally.
• If the REMOTE connector is inserted/extracted with the POWER switch on,
“REMOTE” flashes on the LCD. The tester enters the PROTECTION status
(“PROTECTION” lights up) and cuts off the high-voltage output.
• In making system settings, when FAIL MODE is set to ON, the FAIL or PRO-
TECTION status cannot be cancelled even if the STOP signal is input from the
REMOTE terminal. To cancel the statuses, press the STOP switch on the panel.
For the system settings, see "5.8 System Settings".
NOTE