TOS9213S/ TOS9213AS
6-13
INTERLOCK-signal input requirements
Open: Terminal-to-terminal current of 0.5 mA or less
Short-circuited: Terminal-to-terminal current of 1 mA or more
Connecting the interlock jumper
1.
Insert a scre
w
dri
v
er into A, and open B.
2.
Insert the interlock j
u
mper into B. Take care to pre
v
ent the j
u
mper coat-
ing from
b
ecoming ca
u
ght in B.
3.
Lightly p
u
ll on the j
u
mper to confirm that it is sec
u
rely connected.
4.
Take the same steps for the positi
v
e (+) and negati
v
e (-) terminals.
Then, short-circ
u
it (+) and (-).
To connect a safety device, use the following cable and switch types:
Compatible cables:
• Solid wire: ø0.65 (AWG22)
• Stranded wire: 0.32 mm
2
(AWG22); element-wire diameter: ø0.18 or more
• Removed coating: 11 mm
Rated switch voltage and current:
• 30
V
DC, 0.1 A or more
Circuit for reference
When the door is opened, its contacts open, thus setting the INTERLOCK signal
high. This activates the interlock function.
Connection to the door open/close s
w
itch
A
B
Shank diameter: ø3; tip width: 2.6 mm
11 mm
Door open/close switch