7-18
TOS9213S/ TOS9213AS
DCW:TIMER (D:TIM)
Sets the test time (TEST TIME) and ON/OFF of the timer function for an DC with-
standing voltage test. Also inquires about the current test time and the ON/OFF sta-
tus of the timer function.
Message
Setting:
DCW:TIMER_<
test time
, {ON|OFF}>
D:TIM_<
test time
, {ON|OFF}>
Query:
DCW:TIMER?
D:TIM?
Program data
<test time>
Data format: Real number
Set value:
0.3 to 999
Resolution:
0.1 for 0.3 to 99.9, 1 for 100 to 999
Unit:
s
Program data
<{ON|OFF}>
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the test time to 5 s,
D:TIM 5,ON
Response
To D:TIM
?
, the current test time and the ON/OFF status of the timer function are
returned as “test time, 1/0.”
(Example) If the test time is currently set to 2.0 s and the timer function to OFF,
2.0,0
is returned.
DCW:STARTV (D:STAR)
Sets the start voltage for a DC withstanding voltage test. Also inquires about the set
or setting start voltage.
Message
Setting:
DCW:STARTV_<
start voltage
>
D:STAR_<
start voltage
>
Query:
DCW:STARTV?
D:STAR?
Program data
Data format: Integer
Set value:
0 to 99
Resolution:
1
Unit:
%
(Example) To set the start voltage to 10% of the test voltage,
D:STAR 10
Response
To D:STAR
?
, the current start voltage is returned.
(Example) If the start voltage is currently set to 0%,
0
is returned.