TOS9213S/ TOS9213AS
■
Do not use the tester in locations near a sensitive measuring
instrument or receiver.
Operation in a location subject, may cause such equipment may be affected by noise gener-
ated by the tester.
At a test voltage exceeding 3 k
V
, corona discharge may be gener-
ated to produce substantial amounts of RF broadband emissions between grips on
the test leadwire. To minimize this effect, secure a sufficient distance between alli-
gator clips.
In addition, keep the alligator clips and test leadwire away from the surfaces of con-
ductors (particularly sharp metal ends).
■
Use the product in an industrial environment.
This product may cause interference if used in residential areas. Such use must be
avoided unless the user takes special measures to reduce electromagnetic emissions
to prevent interference to the reception of radio and television broadcasts.
2.3
Precautions for Moving
When moving the tester to the installation site or otherwise transporting it, take the
following precautions:
■
Before moving the tester, turn off the power switch.
Transporting the tester with its POWER switch on can lead to electric shock and
damage.
■
When moving the tester, Disconnect all wires from it.
Moving the tester without disconnecting the cables may result in breakage of the
wire or injury due to the tester tipping over.
■
For transportation, use the special packing material for the
tester.
Transport the tester in its original package to prevent vibration and falls, which may
damage the tester. If you require packing material, contact Kikusui distributor/agent.
■
Be sure to include this manual.