TOS9213S/ TOS9213AS
5-45
Turning the FAIL mode ON/OFF (FAIL MODE)
When the FAIL mode is on, it is impossible to cancel a FAIL judgement or the pro-
tection status using the STOP signal by remote control.
To use the optional high-voltage test probe (HP01-TOS, HP02-TOS), turn on the
FAIL mode. When the test ends with a FAIL judgement, the tester’s FAIL status is
not cancelled even if you release your hand from the probe. To cancel the FAIL sta-
tus, press the STOP switch on the front panel.
The FAIL mode can be set using the SHIFT + F2 keys, regardless of the location of
the cursor. Each time the SHIFT + F2 keys are pressed, ON and OFF alternate. The
mark on the screen is to the left of FAIL MODE.
When the
mark is to the left of MOMENTARY, settings can be made using the
rotary knob.
Turn the rotary knob clockwise
: ON
Turn the rotary knob counterclockwise : OFF
Turning the double-action function ON/OFF (DOUBLE ACTION)
When the double-action function is set to ON, a test cannot be started unless the
START switch is pressed within approximately 0.5 s after the STOP switch is
pressed (“READY” disappears approximately 0.5 s after the STOP switch is
pressed).
To start a test, not only the START switch but also the STOP switch must be
pressed. This bothersome operation is intended to enhance safety.
To control the tester using the GPIB or RS-232C, turn off this function.
The double-action function can be set using the SHIFT + F3 keys, regardless of the
location of the cursor. Each time the SHIFT + F3 keys are pressed, ON and OFF
alternate. The mark on the screen appears to the left of DOUBLE ACTION.
When the mark is to the left of DOUBLE ACTION, settings can be made using
the rotary knob.
Turn the rotary knob clockwise
: ON
Turn the rotary knob counterclockwise : OFF
5.8.2
SYSTEM2
BUZZER
CONTRAST DISCH
SYSTEM 2
SYSTEM 2
BUZZER VOL.
: 4
CONTRAST
: 6
DISCH MODE
: NORM
F1 F2 F3 F4 F5