TOS9213S/ TOS9213AS
5-51
2.
Press the SHIFT + RECALL/STORE keys to display “STO 00:--U
N
TI-
TLTED--” to the right of the title.
3.
Using the rotary kno
b
, set a memory n
u
m
b
er (00 to 99).
4.
Using the key, mo
v
e the c
u
rsor to “--U
N
TITLTED--.”
5.
Using the rotary kno
b
, enter a memory name (
u
sing
u
p to 12 charac-
ters).
Use ASCII-code characters from 20H through 7EH (see the Appendix"A.2
ASCII Code 20H to 7EH").
6.
Press the E
N
TER key to store the test conditions
u
nder the set memory
n
u
m
b
er.
Upon completion of the storage procedure, “STO” changes to “MEM.”
To cancel the storage operation, move the cursor to any location on the display
before pressing the ENTER key.
5.10.2 Recalling panel memory
1.
Press the RECALL key to display “RCL Memory
N
o.: Memory
N
ame” to
the right of the title.
2.
Using the rotary kno
b
, specify the memory n
u
m
b
er to
b
e recalled (00 to
99).
3.
Press the E
N
TER key to recall the test conditions
u
nder the memory
n
u
m
b
er.
“RCL” changes to “MEM” to the right of the title.
To cancel the recall operation, move the cursor to any location on the display before
pressing the ENTER key. Once the recalled test conditions have been changed, the
memory number disappears. The memory number does not reappear even if the pre-
vious test conditions are restored.
VOLT LOWER UPPER
TIMER
2.50
k V
5.00
mA
3.0
s
DCW1
DCW1
LOWER
OFF
TIMER
OFF
UPPER
STO
05 : DCW-TEST5
VOLT LOWER UPPER
TIMER
5.00
k V
5.00
mA
5.0
s
DCW1
DCW1
LOWER
OFF
TIMER
OFF
UPPER
RCL
99 : DCW-TEST99