7-16
TOS9213S/ TOS9213AS
7.4.2
Messages used exclusively for DC withstanding volt-
age testing
This section explains the messages used to check the test conditions and settings for
DC withstanding voltage testing.
DCW:TESTV (D:TES)
Sets the test voltage for a DC withstanding voltage test. Also inquires about the set
or setting test voltage.
• This message is valid even during a test. However, make only fine voltage adjust-
ments rather than large changes to the voltage. A substantial change may activate
the protective circuit to shift the tester to the protection status.
Message
Setting:
DCW:TESTV_<
voltage
>
D:TES_<
voltage
>
Query:
DCW:TESTV?
D:TES?
Program data
Data format: Real number
Set value:
0.00 E3 to 10.5 E3 (or 0 to 10500)
Resolution:
0.01 E3
Unit:
V
(Example) To set the test voltage to 5.00 k
V
,
D:TES 5.00E3
Response
To D:TES
?
, the currently set test voltage is returned.
(Example) If the test voltage is currently set to 2.50 k
V
,
2.50E3
is returned.
DCW:LOWER (D:LOW)
Sets the lower reference (LOWER) and ON/OFF of the lower judgment function for
a DC withstanding voltage test. Also inquires about the lower reference and the ON/
OFF status of the lower judgment function.
Message
Setting:
DCW:LOWER_<
lower reference
, {ON|OFF}>
D:LOW_<
lower reference
, {ON|OFF}>
Query:
DCW:LOWER?
D:LOW?
Program data
<
lower reference
>
Data format: Real number
Set value:
0.01 E-3 to 5.50 E-3
NOTE