8-2
TOS9213S/ TOS9213AS
8.1
Pre-Control Preparation
8.1.1
Connection and startup procedure
1.
T
u
rn off the PO
W
ER s
w
itch on the tester and the TOS6200.
2.
Using an RS-232C ca
b
le, connect the RS-232C connectors on the rear
panel of the tester and the TOS6200.
Use a cross cable of the 9-pin D-SUB female-female AT type as the RS-232C
cable.
3.
T
u
rn on the PO
W
ER s
w
itch on the tester and the TOS6200.
8.1.2
Settings on the TOS6200
Interface settings
On the TOS6200, press the SHIFT + SYSTEM /I/F keys. The LED on the SYSTEM
/I/F key will then light up, and the Interface Settings screen (INTERFACE) will
appear.
On this screen, make the settings specified below.
SPEED (Communication speed) : 19200 bps
DATA (data length)
: 8 bits
PARITY
: NONE
STOP (stop bit)
: 2 bits
Upon completion of the above settings, turn the POWER switch off and then on
again.
For details on the setting procedure, see the Operation Manual for the TOS6200.
Setting the PASS hold time
On the TOS6200, press the SYSTEM key. The LED lights up on the SYSTEM key,
and the System Settings screen (SYSTEM 1) appears.
The PASS hold time must be set to HOLD so that the tester can catch PASS signals
from the TOS6200 without failure.
When the tester recognizes the PASS signal from the TOS6200, the tester enters the
INTER
V
AL mode, regardless of the pass hold time settings.
For details on the setting procedure, see the Operation Manual for the TOS6200.
Setting test conditions
It is necessary to preset test conditions for the TOS6200 in panel memory.
In the TOS6200, memories 1 through 18 contain factory settings that conform to the
relevant safety standards. To add new test conditions, use memories 19 through 99.