4-2
TOS9213S/ TOS9213AS
4.1
Front Panel
Fig.4-1
Front panel
[1] POWER
Used to turn the power on/off. When the power is turned ON ( | ), the tester starts
under the same test conditions as when the power was turned off ( ) at the end of
the preceding test. To start using the factory settings, press the SHIFT key + the
POWER switch to initialize the settings. For details, see "5.15 Initialization".
• With initialization, all contents of the panel memory and stored programs are
cleared. Before starting initialization, confirm that no necessary data remains in
the memory.
[2] STOP
Press this switch under the circumstances listed below. After this switch is pressed,
the tester enters the READY status.
• When you want to stop testing.
• When you want to clear a PASS or FAIL judgement or the PROTECTION
status.
• When you want to stop the forced discharge that occurs after testing (when
the discharge time is set to a value other than 0.0 s).
[3] START
Used to start a test.
The test starts when this switch is pressed while “READY” is displayed on the
LCD.
Once the test starts, the LCD displays “TEST” and the TEST LED flashes on the
indicator during a voltage rise.
When the tester reaches the test voltage, the TEST LED remains lit.
OUTPUT
AUTO
RECALL
ENTER
DCW
IR
SYSTEM
LOCAL
TEST
FUNCTION
GPIB
PASS
FAIL
MEMORY
ENTRY
CURSOR
REMOTE
PROTECTION
RMT
F
F5
F4
F3
F2
F1
PREV
NEXT
EDIT
STORE
I/F
KEY LOCK
SHIF
SHIFT
DANGER
DC WITHSTANDING VOLTAGE
/
INSULATION RESISTANCE TESTER
TOS9213S
HIGH VOLTAGE
TERMINAL
LOW
MAX 10kV
[8] LCD
[5] Indicator
[1] POWER
[2] STOP
[3] START
[17] Rotary knob
[15] SHIFT
[6] FUNCTION
[7] MEMORY
[4] LOCAL/KEYLOCK
[16] CURSOR
[14] F1 to F5
[13] REMOTE terminal
[9] Analog voltmeter
[11] DANGER lamp
[10] Analog-voltmeter
zero adjuster
[12] LOW/GUARD
NOTE