Manufacturer’s declaration-electromagnetic immunity
The Kinsa Smart Ear Thermometer is intended for use in the electromagnetic environment (for home healthcare) specified
below.
The customer or the user of the Kinsa Smart Ear Thermometer should assure that it is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
(for home healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
:
0,15 MHz – 80 MHz
6 Vrms
:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Not applicable
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the
Kinsa Smart Ear Thermometer
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1,2
d = 1,2
80MHz to 800 MHz
d = 2,3
800MHz to 2,7 GHz
Where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.