MKIII OPERATOR’S MANUAL
47
UL
UV Lamp Scan
UL is a test command that results in a UV scan being performed with a Lamp (rather than
the sky) being the source of radiation.
The zenith prism is rotated to the UV dome, FW#2 is set to the 1 position, FW#1 is set to the
3 position, the iris is opened, and the tracker is rotated to the sun. The operator is asked for
lamp number, and lamp-diffuser separation, and the radiation intensity is measured in 1.5nm
increments over the UVB range.
Data is stored in a ULJJJYY.nnn file.
See also Appendix A.
UV
UV(B) Observation
UV results in the irradiance over the “B” region of the spectrum to be measured.
The zenith prism is rotated to the UV dome, FW#2 is set to the 1 position, FW#1 is set to the
3 position, the iris is opened, and the tracker is pointed toward the sun. The UV spectrum is
then scanned in steps of 0.5nm from 290nm to 325nm, and then back to 290 nm.
The data is appended to a UVJJJYY.nnn data file.
When scanning is complete, a calculation of the UVB/UVA McKinley- Diffey weighted
irradiance is computed, sent to the printer and to the monitor screen, and appended to the
DUVJJJYY.nnn data file. A correction is made to the Diffey action spectrum to include the
effects of the UVA region 325-400nm).
UVSUM UV Data Summary
This command initiates activity which processes and prints the daily summary of all UV
measurements made during that day. DUV values are also calculated for each measurement
and data is appended to the UVOAVG.nnn file.
See also Appendix A.
UX
Extended UV Wavelength Scan
UX causes the irradiance over the 286.5nm to 363nm range to be measured in 0.5nm
increments in ascending wavelength. The zenith prism is rotated to the UV dome, FW#2 is
set to the 1 position, FW#1 is set to the 0 position, the iris is opened, and the tracker is
pointed toward the sun. The UV spectrum is scanned, and the data is appended to a
UVJJJYY.nnn data file When scanning is complete, a calculation of the UVB / UVA
McKinley- Diffey weighted irradiance is computed, output to the printer and to the monitor
screen, and appended to the DUVJJJYY.nnn data file.
See also Appendix A.
W0-W4 Time delays
These five commands result in time delays of 1, 5, 10, 20, and 30 minutes respectively and
can be used in command strings or in schedules.
XL
Extended External Lamp Scan
XL is a test command that results in an extended UV scan being performed with a Lamp
(rather than the sky) as the source of radiation.
The zenith prism is rotated to the UV dome, FW#2 is set to the 1 position, FW#1 is set to the
3 position, the iris is opened, and the Tracker is rotated to the sun. The operator is asked for
lamp number and lamp-diffuser separation, and the radiation intensity is measured in 0.5nm
increments over the extended UV range. Data is sampled for 30 cycle time increments
through slit #1 for wavelengths less than 300nm and 20 cycle time increments through slit #5
for higher wavelengths.
Data is stored in an XLJJJYY.nnn file and is normalised to 1 cycle observations.
See also Appendix A.
4. BREWER COMMANDS
Summary of Contents for Brewer Mk III
Page 1: ...Operators manual Spectrophotometer Brewer MkIII...
Page 3: ...ii...
Page 5: ...2...
Page 7: ...4...
Page 25: ...MKIII OPERATOR S MANUAL 22...
Page 29: ...MKIII OPERATOR S MANUAL 26 Figure 3 4 Brewer Spectrophotometer Tripod...
Page 63: ...MKIII OPERATOR S MANUAL 60...
Page 83: ...MKIII OPERATOR S MANUAL 80...
Page 91: ...MKIII OPERATOR S MANUAL 88...
Page 93: ...MKIII OPERATOR S MANUAL 90...
Page 97: ...MKIII OPERATOR S MANUAL 94...
Page 129: ...MKIII OPERATOR S MANUAL 126...
Page 135: ...MKIII OPERATOR S MANUAL 132...