MKIV OPERATOR’S MANUAL
121
19. OPERATIONAL INFORMATION
A variety of operational status values define the current state of the Brewer. Commands exist
to set and read each of these. They share a common syntax so the complete set are listed in
the table below:
Syntax
?<name>
or
?<name>[<index>]
to read the status variable
!<name> <value>
or
!<name>[<index>] <value>
to write the status variable
Response
<value>
? form only
Parameters
parameter / response
format
Meaning
<name>
identifier
parameter in question
<index>
identifier or
number
which of many
<value>
depends on
<name>
depends on <name>
Possible Names
<name>
Format of <value> writeable?
meaning of <value>
ANALOG.NOW
[<analogPt>]
10-bit integer
no
A/D output value in A/D units
BREAK.ABORT.TIME
float seconds
Yes, reset to 0.25
on tepid resets
The elapsed time in seconds for a
continuous TTY line-break to abort
command processing.
BREAK.RESET.TIME
float seconds
Yes, reset to 5.0
on tepid resets
The elapsed time in seconds for a
continuous TTY line-break to force a warm
restart of the Brewer.
BYTE.X [<address>]
ANSI C integer format
(Hex on output)
yes
value of the relevant byte in external RAM
BYTE.F [<address>]
ANSI C integer format
(Hex on output)
no
reads value from Flash memory
BYTE.C [<address>]
ANSI C integer format
(Hex on output)
no
value of the relevant byte in CODE space
BYTE.D [<address>]
ANSI C integer format
(Hex on output)
yes
value of the relevant byte in DATA space
DIGITAL.INPUT
[<digitalPt>]
OFF or ON
no
reads whether input is ON or OFF; logic
levels are transparent to the user.
DIGITAL.OUTPUT
[<digitalPt >]
OFF or ON
yes
digital output point ON or OFF with logic
levels transparent to the user.
ECHO.SUPPRESSION
ON/OFF
Suppresses TTY mode echoing when ON,
allows echoing when OFF.
HG.SWITCH
OFF or ON
yes
mercury lamp set OFF or ON
STD.SWITCH
OFF or ON
yes
standard lamp set OFF or ON
LAMP.POWER [<lamp>] watts
no
lamp power consumption
LAMP.STATE [<lamp>]
lamp state type
yes
state of lamp (OFF or ON)
MOISTURE
Grams/cu. m
no
Moisture content of air inside the Brewer
MOISTURE.HIGH
Grams/cu m
yes
Moisture content high (watermark)
MOTOR.ALLSTILL
TRUE or FALSE
no
True only if all the motors have stopped
moving
APPENDIX H COMPUTER / BREWER INTERFACE (TELETYPE)
Summary of Contents for BREWER MK IV
Page 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Page 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Page 3: ...ii...
Page 5: ...2...
Page 7: ...4...
Page 25: ...MKIV OPERATOR S MANUAL 22...
Page 57: ...MKIV OPERATOR S MANUAL 54...
Page 61: ...MKIV OPERATOR S MANUAL 58...
Page 81: ...MKIV OPERATOR S MANUAL 78...
Page 91: ...MKIV OPERATOR S MANUAL 88...
Page 93: ...MKIV OPERATOR S MANUAL 90...
Page 127: ...MKIV OPERATOR S MANUAL 124...
Page 133: ...MKIV OPERATOR S MANUAL 130...
Page 135: ...MKIV OPERATOR S MANUAL 132...