MKIV OPERATOR'S MANUAL
10
Figure 2.2: View of Brewer with Covers Removed
2.1
SPECTROPHOTOMETER
Table 2.1:
Spectrophotometer Specifications
Slit wavelengths:
‘ozone’ (nm): 303.2 (Hg slit) , 306.3, 310.1, 313.5, 316.8,
320.1 NO
2
(nm): 426.4, 431.4, 437.3, 442.8, 448.1, 453.2
Mercury-calibration (O
3
mode)
302.15 nm
Resolution
0.6 nm in UV; approx 1nm in visible
Stability
±0.01 nm (over full temperature range)
Precision
0.006 ±0.002 nm
Measurement range
286.5 nm to 363.0 nm (in UV);
420 nm to 510 nm (in visible)
Exit-slit mask cycling
0.12 sec/slit, 1.6 sec for full cycle
O
3
Measurement accuracy
±1% (for direct-sun total ozone)
Ambient Operating temperature
range (still air)
0
°
C to +40
°
C (no heater)
-20C to +40
°
C (with heater option)
-50
°
C to +40
°
C (with complete cold weather kit)
Physical dimensions (external
weatherproof container)
Size: 71 by 50 by 28 cm
Weight: 34 kg
Power requirements
Brewer and Tracker
3 A @ 80 to 140 VAC ( with heater option)
1.5A @ 160 to 264 VAC
47 to 440 Hz
Summary of Contents for BREWER MK IV
Page 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Page 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Page 3: ...ii...
Page 5: ...2...
Page 7: ...4...
Page 25: ...MKIV OPERATOR S MANUAL 22...
Page 57: ...MKIV OPERATOR S MANUAL 54...
Page 61: ...MKIV OPERATOR S MANUAL 58...
Page 81: ...MKIV OPERATOR S MANUAL 78...
Page 91: ...MKIV OPERATOR S MANUAL 88...
Page 93: ...MKIV OPERATOR S MANUAL 90...
Page 127: ...MKIV OPERATOR S MANUAL 124...
Page 133: ...MKIV OPERATOR S MANUAL 130...
Page 135: ...MKIV OPERATOR S MANUAL 132...