Introduction 

 

This manual will show you how to install, use and maintain the CVF4 Ventilation Unit. The CVF4 is meant for 
ventilating solar radiation sensors with the purpose to keep the mounted radiometer window or dome clean 
from dew, precipitation and pollution. It can be used outdoors under all-weather conditions. The CVF4 has a 
built in heater to keep the air just above ambient temperature. 

Reading this entire manual is recommended for a full understanding of this product.. 

The triangle with exclamation mark is intended to alert the user to the presence of important 
installation, operating and maintenance instructions in the literature accompanying the 
instrument. 

This electrical equipment should only be serviced by authorized personnel, meaning people who 

have been trained and designated as “authorized” by

 their employers. 

Note  

 

Useful information for the user 

Both the World Meteorological Organisation (WMO) and the Baseline Surface Radiation Network (BSRN) 
guidelines advise using a ventilator to optimise the performance of radiometers in general. 

 

The use of the CVF4 will: 

 

Increase the maintenance interval. 

 

Increase the accuracy of the measured data (cleaner dome / window) 

 

Reduce the zero offset of the radiometer. 

 

The CVF4 is meant for all CM, CMP, SMP, CG, CGR and CUV radiometers. However with the CMP3 and SMP3 
and CGR3 the ventilation effect will be less due to the larger opening around the dome. The CVF4 is meant to 
run continuously. The heater can either be switched on permanently for cold regions or be switched by a 
data logger or clock to remove dew in the morning and be switched off afterwards. In that case the heater is 
operated from a 2 hours before sunrise to an hour after sunrise. This saves power in situations where power 
is limited such as PV operated sites. 

This manual, together with the instruction sheet, gives information related to installation, maintenance, 
product specifications and troubleshooting of the CVF4. 

If any questions should remain, please feel free to contact your Kipp & Zonen dealer or e-mail: 

info@kippzonen.com

 . 

For information about other Kipp & Zonen products or to check for any update of this manual, go to 

www.kippzonen.com

 . 

2.1.

 

Safety precautions 

Many hazards are associated with installing and maintaining instruments on towers or elevated structures. 
It is advised to use qualified personnel for installation and maintenance. The client is responsible for 
following the local safety regulations. 

The use of appropriate equipment and safety practices is mandatory. Check your company's safety 
procedure and protective equipment prior to performing any work. ) If the DustIQ is mounted at a high 
position, special care must be taken to secure both the person installing it and the instrument from falling 
during installation. 

Summary of Contents for CVF4

Page 1: ...Instruction Manual CVF4 Ventilation Unit V2007 01...

Page 2: ...he product has been installed and used according to the instructions supplied by Kipp Zonen Kipp Zonen shall in no event be liable for incidental or consequential damages including without limitation...

Page 3: ...8 4 5 Electrical connections 10 5 Accessories 11 5 1 Spare filters 11 5 2 CMF4 Mounting Fixture 11 5 3 CMB1 Mounting Bracket 11 5 4 Adapter for use of the CVF4 in the CM121C shadow ring 12 6 Maintena...

Page 4: ...eters However with the CMP3 and SMP3 and CGR3 the ventilation effect will be less due to the larger opening around the dome The CVF4 is meant to run continuously The heater can either be switched on p...

Page 5: ...his applies both to the product itself and to any accessories marked with the same symbol Disposal of any such item as unsorted domestic waste is not allowed 2 3 Customer support In case you need supp...

Page 6: ...g holes CMP3 SMP3 and CGR3 4 Mounting holes all other instruments 5 Connector 6 Cover nuts 7 Slit for radiometer cable 3 2 CVF4 bottom side 1 Filter cover 2 Connector box 3 Mounting feet for SOLYS2 Fe...

Page 7: ...washers and Allen key 4 Instruction sheet 5 Pack of 5 spare ventilation fan inlet filters For optional accessories see chapter 3 4 2 Tools required If the CVF4 is mounted with the supplied screws onl...

Page 8: ...ure of the CVF4 on the SOLYS2 Sun Tracker on the next page The locations of the CVF4 mounting holes on the SOLYS2 top mounting plate are shown below CVF4 mounting holes When using your own mounting pl...

Page 9: ...and radiometer levelling might need re adjustment After all mounting screws have been secured the cables of the CVF4 and the radiometer can be connected The radiometer cable goes down through the sli...

Page 10: ...nal It can be connected to a data logger to remotely check proper operation of the CVF4 Ventilator and heater ground are separated but may be connected Ventilator and tacho ground are connected via a...

Page 11: ...mber 2682047 from Kipp Zonen or under EBM Papst part number FP80T from a local distributor of Pabst fans 5 2 CMF4 Mounting Fixture Article number 0362703 5 3 CMB1 Mounting Bracket Article number 03697...

Page 12: ...Adapter for use of the CVF4 in the CM121C shadow ring CVF4 s Remove CVF4s feet and Remount two feet in these counterbore holes Relocation of 2 of the CVF4s mounting feet to fit it on the CM121C shado...

Page 13: ...adapter part is included with the CM121C Note CM121B Article number 0346900 is not supplied with nor aligned for the CVF4 adapter Article number 0346700 CM121C Article number 0346901 is aligned for a...

Page 14: ...output is logged with the radiometer data the performance of the ventilator can be remotely checked 1 Refer to radiometer instruction sheet for maintenance instructions 2 Regularly inspect the fan inl...

Page 15: ...le 25 m Temperature range 40 C to 70 C Increase of air temperature 0 25 K without heating 1 K at 6 8 W heating Compliance with standards ISO TR 9901 BSRN operations manual Zero o set A reduction 50 re...

Page 16: ...en product please contact your local representative in the first instance The information can be found in the Contact section of our website at www kippzonen com 1 Alternatively you can contact us dir...