INSTALLATION AND MANTENANCE
Instruction manual PAR-LITE
21
4.1 Recalibration
Recalibration is suggested every two years. This can be done
in two ways. The first is by comparing with the measurement
of a similar sensor at the same site. Preferably daily totals of
several days (clear and overcast) should be compared.
Calibration factor could be corrected if results differ by more
than five percent.
The second way is to let a recalibration be performed at the
PAR
LITE
factory.
4.2 Adjustment of calibration
If necessary, the sensitivity of the PAR
LITE
can be adapted.
This can be done by soldering a resistor between the + (white)
and - (green) output wires. In this way the internal resistance
is shunted. Suppose, the internal (load) resistance R
l
is 270
Ω
.
To be sure, measure the impedance of the PAR
LITE
with an
Ohmmeter, in the correct direction. (reverse voltage on
photodiode). Neglecting the cable-resistance, the cable
resistance is 0.12
Ω
per meter a shuntresistor R
p
of 2430
Ω
reduce the sensitivity by a factor of 0.9. A shunt resistor of
about 2430
Ω
can be made with a 2700
Ω
resistor shunted by
a resistor of 24 k
Ω
The general formula is: S
new
= S
old
.R
p
/(R
l
+ R
p
)
Summary of Contents for PAR lite
Page 1: ...Instruction Manual Photosynthetic Active Radiometer PAR lite ...
Page 6: ...CONTENTS Instruction manual PAR LITE 3 ...
Page 22: ...CALIBRATION Instruction manual PAR LITE 19 ...
Page 26: ...INSTALLATION AND MANTENANCE Instruction manual PAR LITE 23 ...
Page 30: ...DELIVERY Instruction manual PAR LITE 27 ...
Page 31: ...ACCESSORIES Instruction manual PAR LITE 28 7 Accessories CLF1 levelling fixture 0338700 ...