Consequences of modified target temperature
As soon as the target temperature is reached, the
temperature warning limits must be adjusted so
that the target temperature is above or below the
temperature warning limits. Otherwise the tem-
perature alarm will be triggered
“Displaying and changing the temperature warning
The set target temperature is saved automatically.
The target temperature is retained after a power
failure and after the unit has been switched off.
7.
When the unit has reached its target temperature, stock the
unit.
6.2.2 Temperature warning limits
6.2.2.1
Function of the temperature warning limits
The temperature warning limits define how much deviation the unit
will tolerate between the actual temperature and the target tem-
perature. The temperature warning limits are unit-specific. They
can be adapted to the requirements of the chilled goods.
The values for the temperature warning limits are set at the factory
and comply with the valid DIN standards for the unit.
Recommended temperature warning limits
The temperature warning limits must not be the
same as the target temperature.
For blood reserves, set the temperature limits as
follows:
–
Upper temperature warning limit: at least 2 °C
higher than the target temperature
–
Lower temperature warning limit: at least 2 °C
lower than the target temperature
Recommended temperature warning limits for
FROSTER
To ensure that the unit functions properly, the tem-
perature warning limits must be set for FROSTER
models as shown in table
Commissioning
Programming the unit > Temperature warning limits